The detector for an HPLC is the component that emits a response due to the eluting sample compound and subsequently signals a peak on the chromatogram. It is positioned immediately posterior to the stationary phase in order to detect the compounds as they elute from the column. The bandwidth and height of the peaks may usually be adjusted using the coarse and fine tuning controls, and the detection and sensitivity parameters may also be controlled (in most cases). There are many types of detectors that can be used with HPLC. Some of the more common detectors include: Refractive Index (RI), Ultra-Violet (UV), Fluorescent, Radiochemical, Electrochemical, Near-Infra Red (Near-IR), Mass Spectroscopy (MS), Nuclear Magnetic Resonance (NMR), and Light Scattering (LS). Refractive Index (RI) detectors measure the ability of sample molecules to bend or refract light. This property for each molecule or compound is called its refractive index. For most RI detectors, light proceeds through a bi-modular flow-cell to a photodetector. One channel of the flow-cell directs the mobile phase passing through the column while the other directs only the mobile phase. Detection occurs when the light is bent due to samples eluting from the column, and this is read as a disparity between the two channels.

Ultra-Violet (UV) detectors measure the ability of a sample to absorb light. This can be accomplished at one or several wavelengths:

A) Fixed Wavelength measures at one wavelength, usually 254 nm

B) Variable Wavelength measures at one wavelength at a time, but can detect over a wide range of wavelenths

C) Diode Array measures a spectrum of wavelengths simulateneously

UV detectors have a sensitivity to approximately 10-8 or 10 -9 gm/ml.

Fluorescent detectors measure the ability of a compound to absorb then re-emit light at given wavelengths. Each compound has a characteristic fluorescence. The excitation source passes through the flow-cell to a photodetector while a monochromator measures the emission wavelengths.

Has sensitivity limit of 10-9 to 10-11 gm/ml.

Radiochemical detection involves the use of radiolabeled material, usually tritium (3H) or carbon-14 (14C). It operates by detection of fluorescence associated with beta-particle ionization, and it is most popular in metabolite research. Two detector types:

A) Homogeneous- Where addition of scintillation fluid to column effluent causes fluorescence.

B) Heterogeneous- Where lithium silicate and fluorescence caused by beta-particle emission interact with the detector cell.

Has sensitivity limit up to 10-9 to 10-10 gm/ml.

Electrochemical detectors measure compounds that undergo oxidation or reduction reactions. Usually accomplished by measuring gain or loss of electrons from migrating samples as they pass between electrodes at a given difference in electrical potential.

Has sensitivity of 10-12 to 10-13 gm/ml

Mass Spectroscopy (MS) Detectors- The sample compound or molecule is ionized, it is passed through a mass analyzer, and the ion current is detected. There are various methods for ionization:

A) Electron Impact (EI)- An electron current or beam created under high electric potential is used to ionize the sample migrating off the column.

B) Chemical Ionization- A less aggresive method which utilizes ionized gas to remove electrons from the compounds eluting from the column.

C) Fast Atom Bombarbment (FAB)- Xenon atoms are propelled at high speed in order to ionize the eluents from the column.

Has detection limit of 10-8 to 10-10 gm/ml.

Nuclear Magnetic Resonance (NMR) Detectors- Certain nuclei with odd- numbered masses, including H and 13C, spin about an axis in a random fashion. However, when placed between poles of a strong magnet, the spins are aligned either parallel or anti-parallel to the magnetic field, with the parallel orientation favored since it is slightly lower in energy. The nuclei are then irradiated with electromagnetic radiation which is absorbed and places the parallel nuclei into a higher energy state; consequently, they are now in "resonance" with the radiation. Each H or C will produce different spectra depending on their location and adjacent molecules, or elements in the compound, because all nuclei in molecules are surrounded by electron clouds which change the encompassing magnetic field and thereby alter the absorption frequency.

Light-Scattering (LS) Detectors- When a source emits a parallel beam of light which strikes particles in solution, some light is reflected, absorbed, transmitted, or scattered. Two forms of LS detection may be used to measure the two latter occurrences:

A) Nephelometry- This is defined as the measurement of light scattered by a particulate solution. This method enables the detection of the portion of light scattered at a multitude of angles. The sensitivity depends on the absence of background light or scatter since the detection occurs at a black or null background.

B) Turbidimetry- This is defined as the measure of the reduction of light transmitted due to particles in solution. It measures the light scatter as a decrease in the light that is transmitted through the particulate solution. Therefore, it quantifies the residual light transmitted. Sensitivity of this method depends on the sensitivity of the machine employed, which can range from a simple spectrophotometer to a sophisticated discrete analyzer. Thus, the measurement of a decrease in transmitted light from a large signal of transmitted light is limited to the photometric accuracy and limitations of the instrument employed.

Near-Infrared Detectors- Operates by scanning compounds in a spectrum from 700 to 1100 nm. Stretching and bending vibrations of particular chemical bonds in each molecule are detected at certain wavelengths. This is a method which offers several advantages: speed (sometimes less than 1 second), simplicity of preparation of sample, multiple analyses from single spectrum, and nonconsumption of the sample.

User Avatar

Wiki User

14y ago
This answer is:
User Avatar
More answers
User Avatar

Wiki User

7y ago

The flame ionization detector (FID) is the most commonly used.

This answer is:
User Avatar

Add your answer:

Earn +20 pts
Q: What detectors are used in HPLC?
Write your answer...
Submit
Still have questions?
magnify glass
imp
Related questions

Where can one purchase HPLC detectors?

You can purchase used HPLC detectors and other equipment from the usedhplc website or from the ebay bidding website. Alternatively you can buy HPLC detectors from the equipnet website.


How do you distinguised np-hplc and rp-hplc?

NP-HPLC is "Normal Phase" HPLC, wherein the solvents used are less polar than the substrate in the HPLC column (e.g. using hexane or dichloromethane with a silica HPLC column). RP-HPLC is "Reverse-Phase" HPLC, wherein the solvents used are more polar than the substrate in the HPLC column (e.g. using Water and Methanol with a octadecylsilane (ODS or C18) column).


Definition of standard in HPLC?

Answer: HPLC standards are an indispensible tool for analytical HPLC applications. They are used to monitor column performance & calibrate detector response.


What is difference bet GLC and HPLC?

GLC has a stationary liquid phase and gas moving phase HPLC had a stationary solid phase and liquid moving phase HPLC is done under high pressure. HPLC can be used for thermally unstable compounds as opposed to GLC HPLC can be used for polar or low volatile compounds as opposed to GLC


How does one apply the HPLC princple?

The HPLC principle is used to separate liquid mixtures into individual parts. It's frequently used in several field such as medicine and law. HPLC is specifically used in sports to test for performance enhancing drugs in urine.


What is assay by HPLC?

HPLC is high-performance liquid chromatography. This method is used for the separation and production of chemicals but also for chemical analysis.


Why is Erbium perchlorate is used in calibration of HPLC?

Erbium has a strong absorption in uv and visible range, It is used in HPlc calibration for the wavelength accuracy verification of the PDA detector.


Why propyl paraben used in HPLC calibration?

propyl paraben being a stable compound with reproducible chromatographic parameters used as a internal standard for the calibration of the HPLC systems


What is dead volume HPLC?

The dead volume in HPLC is 137.45. The dead volume in science is used in retention measurements and also in thermodynamic studies and the abbreviation HPLC stands for High Pressure Liquid Chromatography.


Which are the top commercial uses for HPLC?

HPLC is used to analyze, identify, purify and quantify organic compounds.


What was used in fire detectors before americium?

The first smoke detectors are based on optical principles - photoelectric detectors.


Which equipment is used to separate mixtures?

HPLC, HPTLC, GPC